Principle


Dielectric constant (Dk) is a measure of how readily a material absorbs energy while in the presence of a field. Dissipation factor (Df) is a measure of how readily that material gives up the stored energy after the field has been removed.


Capabilities


Capacitance Measurement and via Resonance Cavities


Equipment



Equipment Type

Information

                  

Agilent LCR Meter
  • Frequency range 20Hz - 2MHz
  • Measures Dk as ratio of material to vacuum capacitance
  • Measures Df as ratio of current through resistance and capacitance circuits

               

Agilent Parallel Plate Electrode 16451B

  • Can be tested as contact or gap method
  • Four different electrodes, depending on material type and method

Equipment Type

Information

                      

Agilent VNA 8722C

Custom Rectangular Prism Resonance Cavities

  • Frequency range up to 40GHz
  • Resonance cavities: 1, 1.5, 2, 2.4, 2.5, 3. 5. 6. 10, 20, 30, 40GHz
  • Measures Dk/Df by shift in resonance peal and quality factor (Q)

          

Compass Technology Custom Rectangular Prism Resonance Cavity

  • Valid only for 2.45GHz
  • Measures Dk/Df by shift in resonance peal and quality factor (Q)